Join our host and Inorganic Ventures Technical Director, Mike Booth, as he sparks insightful conversations with our panel of experts in Bench Boost, your go-to podcast for analytical chemistry enthusiasts. Together, they explore the realm of ICP, sharing tips and tricks and diving deep into all things analytical chemistry.
Bench Boost by Inorganic Ventures
CCVs & ICVs
Use Left/Right to seek, Home/End to jump to start or end. Hold shift to jump forward or backward.
This week on Bench Boost Mike explains two key ICP-OES/ICP-MS quality control samples: the Initial Calibration Verification (ICV) and Continuing Calibration Verification (CCV). The ICV is run after calibration and before samples to confirm the calibration worked. The CCV is run periodically during the batch to confirm the calibration remains valid as drift, clogging, cone buildup, washout, or other issues occur. This episode concludes with reviewing how to troubleshoot ICV or CCV failures.