Bench Boost by Inorganic Ventures
Join our host and Inorganic Ventures Technical Director, Mike Booth, as he sparks insightful conversations with our panel of experts in Bench Boost, your go-to podcast for analytical chemistry enthusiasts. Together, they explore the realm of ICP, sharing tips and tricks and diving deep into all things analytical chemistry.
Bench Boost by Inorganic Ventures
pH and USP 791
Use Left/Right to seek, Home/End to jump to start or end. Hold shift to jump forward or backward.
This week on Bench Boost Mike explains why accurate pH measurement is more complex than it appears, highlighting the effects from temperature, ionic strength, calibration technique, probe condition, and sample chemistry. He reviews pH theory of hydrogen ion activity (not just concentration) and the logarithmic meaning of pH changes, then describes how a pH probe functions as an electrochemical cell. He details temperature impacts on solution pH and electrode response (Nernst slope), notes automatic temperature compensation limits, summarizes USP <791> calibration buffers and allowable uncertainty.